Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Medtronic · MMT- 508 - MiniMed pump

Official description · CVE.org

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Breakdown
14.4 / 100
Exploited
Not confirmed
Reports
0
Tactics
3/15

How an attack unfolds

Based on mappings. Select a step to see the evidence.
  1. How attackers get in

  2. What they do

Disclosed on 13 Aug 2018. No public exploit code or confirmed exploitation is recorded.

techniques

2 , grouped by in kill-chain order 2

Initial AccessTA0001

MITRE CTID mapping of the CVE · primary impact

Credential AccessTA0006

Network Sniffing T1040 also Discovery
MITRE CTID mapping of the CVE · how it is exploited

Detection

0 rules for this CVE · 158 general rules for its techniques

Rules for this CVE

0 that reference CVE-2018-14781 directly.

CVE-specific rules are not available in this build. Search SigmaHQ ↗

Sigma rules © SigmaHQ, Detection Rule License 1.1; linked, not reproduced.

What security researchers reported

Not named in any threat report yet

References

3 links, grouped by NVD tag