Patchfirst
CVE-2019-6538 Critical · 9.3 EPSS 0.9%

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

Medtronic · Conexus Radio Frequency Telemetry Protocol

Summary

The Conexus Radio Frequency Telemetry Protocol used across numerous Medtronic implantable cardiac devices and their associated monitors and programmers, including CareLink and MyCareLink products and a range of ICD and CRT-D models, does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, while its radio is enabled and without credentials or user interaction, can inject, replay, modify, or intercept telemetry data and use the protocol's memory read/write capability to alter memory in an implanted cardiac device.

Official description · CVE.org

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Breakdown
19.8 / 100
Exploited
Not confirmed
Reports
0
Tactics
5/15

How an attack unfolds

Based on mappings. Select a step to see the evidence.
  1. What they do

Disclosed on 25 Mar 2019. No public exploit code or confirmed exploitation is recorded.

techniques

4 , grouped by in kill-chain order 4

ExecutionTA0002

Hijack Execution Flow T1574 also Stealth
MITRE CTID mapping of the CVE · primary impact

Credential AccessTA0006

Adversary-in-the-Middle T1557 also Collection
MITRE CTID mapping of the CVE · primary impact

CollectionTA0009

MITRE CTID mapping of the CVE · primary impact

ImpactTA0040

MITRE CTID mapping of the CVE · primary impact

Detection

0 rules for this CVE · 35 general rules for its techniques

Rules for this CVE

0 that reference CVE-2019-6538 directly.

CVE-specific rules are not available in this build. Search SigmaHQ ↗

General rules for this CVE’s techniques

These detect the technique in general, not this specific vulnerability.

Hijack Execution FlowT15748 general rules
Adversary-in-the-MiddleT155710 general rules
Data from Local SystemT100514 general rules
Application or System ExploitationT1499.0043 general rules

Sigma rules © SigmaHQ, Detection Rule License 1.1; linked, not reproduced.

What security researchers reported

Not named in any threat report yet

Also mentioned in 3 community threat pulses (AlienVault OTX)

Weaker evidence

Pulses are indicator collections that anyone can publish. A mention is not confirmation of exploitation.

References

2 links, grouped by NVD tag
↑ ↓ chooseEnter openEsc close